Sviluppo di un tomografo multi-energy per lo studio pre-clinico di nuove metodiche diagnostiche finalizzate al riconoscimento precoce della patologia tumorale

Masetti, Simone (2008) Sviluppo di un tomografo multi-energy per lo studio pre-clinico di nuove metodiche diagnostiche finalizzate al riconoscimento precoce della patologia tumorale, [Dissertation thesis], Alma Mater Studiorum Università di Bologna. Dottorato di ricerca in Fisica, 20 Ciclo. DOI 10.6092/unibo/amsdottorato/841.
Documenti full-text disponibili:
[img]
Anteprima
Documento PDF (Italiano) - Richiede un lettore di PDF come Xpdf o Adobe Acrobat Reader
Download (14MB) | Anteprima

Abstract

A new multi-energy CT for small animals is being developed at the Physics Department of the University of Bologna, Italy. The system makes use of a set of quasi-monochromatic X-ray beams, with energy tunable in a range from 26 KeV to 72 KeV. These beams are produced by Bragg diffraction on a Highly Oriented Pyrolytic Graphite crystal. With quasi-monochromatic sources it is possible to perform multi-energy investigation in a more effective way, as compared with conventional X-ray tubes. Multi-energy techniques allow extracting physical information from the materials, such as effective atomic number, mass-thickness, density, that can be used to distinguish and quantitatively characterize the irradiated tissues. The aim of the system is the investigation and the development of new pre-clinic methods for the early detection of the tumors in small animals. An innovative technique, the Triple-Energy Radiography with Contrast Medium (TER), has been successfully implemented on our system. TER consist in combining a set of three quasi-monochromatic images of an object, in order to obtain a corresponding set of three single-tissue images, which are the mass-thickness map of three reference materials. TER can be applied to the quantitative mass-thickness-map reconstruction of a contrast medium, because it is able to remove completely the signal due to other tissues (i.e. the structural background noise). The technique is very sensitive to the contrast medium and is insensitive to the superposition of different materials. The method is a good candidate to the early detection of the tumor angiogenesis in mice. In this work we describe the tomographic system, with a particular focus on the quasi-monochromatic source. Moreover the TER method is presented with some preliminary results about small animal imaging.

Abstract
Tipologia del documento
Tesi di dottorato
Autore
Masetti, Simone
Supervisore
Dottorato di ricerca
Ciclo
20
Coordinatore
Settore disciplinare
Settore concorsuale
Parole chiave
multi-energy tomografia raggi x monocromatici angiogenesi
URN:NBN
DOI
10.6092/unibo/amsdottorato/841
Data di discussione
12 Giugno 2008
URI

Altri metadati

Statistica sui download

Gestione del documento: Visualizza la tesi

^